Force and mechanical measuring equipment
– Measure pH/Conductivity (EC)/Total Dissolved Solids (TDS) Simultaneously.
+ HI98129: Conductivity follows µS/cm and TDS in mg/L.
+ HI98130: Conductivity follows mS/cm and TDS in g/L.
– Compact, pocket-friendly design.
– The pH electrode is removable and replaceable.
– Automatic temperature compensation.
– Provide a complete set including meter, electrode and standard solution, washing and storing electrodes.
REMOVABLE pH ELECTRODES
HI98129 waterproof, highly accurate pH meter can measure EC/TDS with fabric joints. Expanded fabric joints increase electrode life and the pH electrode is replaceable. Additionally, the replacement electrode has no pins so it won't bend or break.
EC/TDS GRAPHIC DETECTOR
The conductive graphite probe provides greater accuracy because it is not contaminated by salts in the solution.
AUTOMATIC TEMPERATURE COMPENSATION
The machine integrates a temperature sensor for quick response time and ensures temperature compensation for highly accurate measurement results.
AUTOMATIC CALIBRATION
The machine has an automatic calibration function with supplied calibration solutions
– pH: Calibration at 2 points at pH 4.01 / 7.01 / 10.01
– EC/TDS: 1 point calibration at 1413 µS/cm, 1382 ppm (mg/L)
Because EC and TDS are correlated, you only need to calibrate 1 EC or TDS point for both parameters
HOLD BUTTON
Helps keep a stable value on the screen to facilitate recording and storing measurement values
STABLE SYMBOL
The hourglass icon will flash during measurement and disappear to indicate that the measurement result has stabilized and this value has been taken.
SUPPLIED COMPLETE SET
The HI98129 is supplied as a complete set to perform measurements without any additional purchase:
– Meter and internal pH, EC and temperature electrodes
– Calibration solutions pH4.01 and 7.01
– EC and TDS calibration solutions
– Electrode cleaning solution after measurement
– Electrode preservation solution.
SPECIFICATIONS
The scale |
pH |
0.00 to 14.00 pH |
EC |
0 to 3999 µS/cm |
|
TDS |
0 to 2000 ppm (mg/L) |
|
To |
0.0 to 60.0°C / 32.0 to 140.0°F |
|
Resolution |
pH |
0.01 pH |
EC |
1 µS/cm |
|
TDS |
1 ppm (mg/L) |
|
To |
0.1°C / 0.1°F |
|
Accuracy |
pH |
±0.05 pH |
EC |
±2% full scale |
|
TDS |
±2% full scale |
|
To |
±0.5°C /±1°F |
|
Calibration |
EC/TDS |
automatic, 1 point at: 1413 µS/cm, 1382 ppm (mg/L) |
pH |
automatic, 2 points at pH 4.01 / 7.01 / 10.01 |
|
Heat compensation |
EC/TDS |
automatic |
pH |
automatic with β from 0.0 to 2.4% /°C |
|
TDS conversion factor |
0.45 to 1.00 |
|
pH electrode |
HI73127 |
|
The battery |
4 x 1.5V |
|
Environment |
0 to 50°C (32 to 122°F); RH max 100% |
|
Size |
163 x 40 x 26 mm |
|
Mass |
100 g |
|
Guarantee |
6 months for the machine and 3 months for the electrodes. (for products with intact tags and warranty card). |
|
Supply includes |
– HI98129 meter – pH electrode HI73127 (in machine) – EC/TDS probe and temperature (fixed in the machine). – Solution package pH standards 4.01 and 7.01 – Solution package standard 1413 µS/cm and 1382 mg/L – Solution package wash electrode – Solution package preserve electrode – Electrode removal tool – Battery (in the device) - User manual - Warranty. – Paper container |