– Ion selective electrode (ISE) half-cell
– Must be used with HI5315 reference electrode
– Solid state sensor
– Measures from 0.026 mg/L to 260 mg/L CN–
Cyanide selective electrode HI4009 is a half-cell sensor housed in an epoxy body. The internal elements are isolated from the sample by a silver halide sensing membrane. This sensor requires a separate reference electrode when measuring.
Hanna solid state half-cell electrodes are designed for use with the HI5315 conical junction reference electrode. The conical reference forms a liquid junction with the sample solution at the tip of the conical junction, providing a very stable reference current and a consistent flow rate of electrolyte into the solution.
For HI4009 To accurately measure iodine content, it is important to The ionic strength of the standard and sample solutions is adjusted to a constant value. with an ionic strength adjustment solution (ISA). Alkaline ISAs ensure that the activity coefficient is constant so that the concentration of free cyanide ions can be measured. Measuring all solutions at a constant ionic strength reduces the margin of error between measurements. HI4009 Cyanide Solid State ISE can measure from 1μM (0.026 mg/L) to 0.01 M (260 mg/L).
Conversion table to Cu2+
|
Multiply by |
mol/L (M) to ppm (mg/L) |
2.602 x 104 |
ppm (mg/L) to M (mol/L) |
3.843 x 10-5 |
SALIENT FEATURES
Fixed sensor module – Solid-state half-cell ISE is the electrode detects free ions in solution. The highly compressed, insoluble sensing particles are attached to the halfcell electrode by a fixed sensing module for ease of use.
Durable epoxy body – Half-cell probe body made of durable epoxy resin.
BNC connection – The HI4009 has a universal BNC connector that easily connects to any meter with a BNC input.
Working principle: A solid-state sensor electrode develops a voltage due to the ion exchange that occurs between the solution and the inorganic membrane. An equilibrium occurs due to the limited solubility of the membrane material in the sample. When the ionic strength of the solution is fixed with ISA SOLUTION, the voltage will be proportional to the concentration of free ions in the solution.
ISE ELECTRODE AND NECESSARY SOLUTIONS AND ACCESSORIES
https://hannavietnam.com/blog-detail/dien-cuc-ise-va-cac-dung-dich-phu-kien-can-thiet-122
SPECIFICATIONS
Type |
– half-cell; solid-state – Use with HI5315 reference electrode (required) |
The scale |
0.01 M to 1×10-6M CN– 0.02 to 260 mg/L (ppm) |
Optimal pH scale |
>11 |
Temperature scale |
0 to 80 °C |
Slope |
-56 mV |
Diameter |
12mm |
Total length |
120 mm |
Body material |
epoxy |
Cable |
Coaxial; 1 m |
Connect |
BNC |
Guarantee |
06 months (For products with intact tags and warranty card) |
Compatible Meters |
– HI98191 portable ISE meter – ISE benchtop meter HI5222, HI5522 – HI2550 benchtop ISE meter |
Used for |
determination of free cyanide ions in plating baths, wastewater and in plant and soil samples |