Semiki logo
HOTLINE: +84 979761016
0

Cart

No products in the cart.

Continue shopping
Home / Shop / Sodium/Sodium Chloride Measuring Electrode FC300B
Product portfolio

Sodium/Sodium Chloride Measuring Electrode FC300B

Producer:
SKU: HANNA-FC300B Category:

The FC300B is an ion selective electrode (ISE) used to determine the sodium ion (Na+) content in solution.

The glass membrane generates an electric current from the exchange of sodium ions between the membrane surface and the sample. The sensor is located inside the glass body.

The FC300B is ideal for laboratory measurements, food and beverage production and water quality assessment.

Electrolytically refillable glass body

Single ceramic joint

Measures from 0.23 to 22990 Na+

YouTube video

The FC300B is a combination silver/silver chloride (Ag/AgCl) reference electrode and a sodium ion selective electrode. Both are housed in a glass body filled with ammonium chloride/silver chloride solution.

The sodium sensitive membrane should be conditioned prior to calibration. Conditioning of the electrode is a two-step process in which the glass is etched and then rehydrated. The first step is etching. This process removes a thin layer of the sodium sensitive glass membrane, exposing a new sodium-containing surface. The second step is rehydration, in which the probe is placed in a storage solution to regenerate the glass membrane. Allowing the glass to regenerate after etching creates optimal conditions for sodium ion exchange between the membrane and the solution.

For the FC300B to measure sodium accurately, it is important that the ionic activity of the standards and samples be adjusted to a high and constant value with a sodium ion activity adjustment (ISA) solution. The sodium ISA is important for two reasons. One is that the ISA buffers the pH of the standards and samples above pH 9.8. This eliminates any interference from hydrogen ions at lower pHs. The second purpose of the ISA is to ensure that the activity coefficient is constant so that the concentration of free sodium ions can be measured. Solutions with a constant ionic strength reduce errors between different samples with different concentrations.

The FC300B sodium glass can measure from 10μM (0.23 mg/L) to 1 M (22,990 mg/L).

NA+ AND NACL CONVERSION TABLE

 

Multiply by

moles/L (M) Na+ to ppm (mg/L) Na+

22990

ppm (mg/L) Na+ to M (moles/L) Na+

4.35 x 10-5

ppm (mg/L) Na+ to ppm (mg/L) NaCl

2.54197

ppm (mg/L) NaCl to ppm (mg/L) Na+

0.39339

 

ISE ELECTRODE AND NECESSARY SOLUTIONS AND ACCESSORIES

https://hannavietnam.com/blog-detail/dien-cuc-ise-va-cac-dung-dich-phu-kien-can-thiet-122

 

 

Specifications

Type

glass, combination

The scale

Sodium (Na+) 1.0M to 1 x 10-5M, 22,990 to 0.23 mg/L (ppm)

Optimal pH

9.75 to 14

Optimal temperature

0 to 80ºC

Slope

+57

Diameter

12mm

Body length

120 mm

Body material

glass

Cable

1m

Connector

BNC

Guarantee

06 months

Request a quote
SEMIKI MEASUREMENT EQUIPMENT COMPANY LIMITED
Headquarters:
  • Floor 12 – Tower A2, Viettel Building, 285 Cach Mang Thang Tam, Ward 12, District 10, Ho Chi Minh City, Vietnam
  • Sales@semiki.com
  • +84 979761016
  • Tax code 0313928935
Warranty, maintenance and repair center:
  • 10th Floor, Halo Building, 19-19/2A Ho Van Hue, Ward 9, Phu Nhuan District, HCMC
Representative office in Hanoi:
  • 9th Floor, 3D Building, No. 3 Duy Tan Street, Dich Vong Hau Ward, Cau Giay District, Hanoi City
© 2024 Semiki inc. All rights reserved.

Ask SEMIKI for a quote

Complete the form below to receive a quote from SEMIKI.
You need to enter all information fields below.