Force and mechanical measuring equipment
– Integrated temperature sensor and temperature compensation
– Flat probe helps reduce dirt sticking to the electrode
– Dedicated glass sensor for fast stabilization and accurate results
– PVDF (Kynar®) body is easy to clean, disinfect, durable, resistant to most chemicals (electronic solvents, sodium hypochlorite)
– 3/4”NPT connection thread
– Maximum pressure 6 bar (87 psi)
– Matching pin helps avoid problems caused by ground loop currents
– Digital probe stores model, firmware, serial number and calibration information
TECHNICAL SPECIFICATIONS OF pH ELECTRODE HI10XX-Y8ZZ
HI10XX-Y8ZZ |
||
In there: |
||
XX = |
06 |
PTFE Joint |
16 |
Ceramic Joint |
|
Y = |
1 |
Low temperature glass sensor LR, titanium pin maching -5 to 80oC (23 to 176oF) Temperature compensation: 0 to 80 oC (32 to 176 oF) 0 to 12 pH |
3 |
HR high temperature glass sensor, titanium pin maching 0 to 100 oC (32 to 212 oF) Temperature compensation: 0 to 100 oC (32 to 176 oF) 0 to 14 pH |
|
4 |
Fluoride-resistant (HF) glass sensor, titanium pin maching -5 to 60 oC (23 to 140 oF) Temperature compensation: 0 to 60 oC (32 to 140oF) 0 to 10 pH |
|
8 |
|
Smart probe, with RS485 connection port |
ZZ |
|
00, 05, 10, 15, 25, 50; cable length (m) |
NOTE:
– Line HI1006-18 and HI1016-18, used in environments with low conductivity or low temperature
– Line HI1006-38 and HI1016-38, used in environments with wide pH range or high temperature
– Line HI1006-48 and HI1016-48, used in environments where there is the presence of hydrofluoric acid
TECHNICAL SPECIFICATIONS OF pH ELECTRODE HI1006-1805
HI1006-1805 |
|
pH scale |
0.00 to 12.00 pH |
pH accuracy |
± 0.02 pH |
Temperature |
-5.0 to 80 oC / 23 to 176 oF |
Temperature accuracy |
± 0.5 oC / 1.0 oF |
Heat compensation |
Automatic or manual from 0.0 to 80.0 oC (32 to 176 oF) |
Body material |
PVDF |
Joint |
PTFE |
Sensor |
LT Low Temperature Glass |
Sensor head |
Plane |
Maximum pressure |
6 bar |
Threaded connection |
3/4″ NPT |
Cable length |
5m |