Force and mechanical measuring equipment
– PTFE joints.
– No batteries required.
– Matching pin.
– Flat electrode tip Minimize breakage due to accidental impact.
– Cable length 10m.
– 3/4″ male thread is popular when installing in pipes.
FLAT ELECTRODE HEAD
– Helps reduce plaque buildup and accidental impact
– Easy to clean periodically.
MATCHING PIN
In many industrial applications, especially plating baths, ground loop currents are a very common problem. When using a traditional electrode/controller system with an electrode reference connected to both the electrode and the instrument, current flows through the reference half cell, causing fluctuations in the reading and serious damage to the Ag/AgCl element. The potential coupling pin shields the reference from external electrical shields. In the figure above, the coupling pin allows for stable measurements and ensures an efficient process. For proper operation, the coupling pin must be continuously immersed in the measured solution and is therefore located close to the electrode junction.
The scale |
0 to 13 pH |
Joint |
PTFE |
Anti-fouling ability |
Not stain resistant. The electrode must be washed and recalibrated at least once a week with clean water or more often depending on the actual water environment because a dirty electrode will be damaged. |
Type |
Flat electrode without battery, no anti-fouling |
Measuring temperature |
From -5 to 80°C – The electrode only measures samples at room temperature. – The electrode cannot withstand high temperatures, the higher the temperature, the shorter the electrode life. |
Maximum pressure |
6 bar |
Connector |
BNC |
Use for |
Meter with BNC connection port |
Cable |
10m |
Guarantee |
6 months (for products with intact seals and warranty cards) *Warranty only for manufacturer defects, because the electrode is not anti-stick and cannot withstand high temperatures, so the warranty is not valid if measuring in high temperatures, strong chemicals or strong corrosion or if stuck due to the measuring environment, or due to not washing the electrode... |
Supply includes |
– HI1006-20 electrode with 10m long cable attached. (The electrode body is printed as HI1006-20, and 10 means the cable is 10m long) – Protective cover when not in use. – Warranty. – Original quality certificate from factory. – Paper container. pH electrodes are NOT resistant to high temperatures and strong corrosive chemicals. |