Force and mechanical measuring equipment
– 5 cuvettes already contain standard solution absorbance at 420nm, 466nm, 525nm, 610nm, and zero.
– Only used for HANNA Photometer HI83300.
– Supplied with a light-proof box, ensuring standard quality.
– Expiry date 1 year.
– Use until expiration date.
– Supplied with COA and MSDS.
LIGHT-SHARDED CONTAINER BOX, AVOID COLLISION
– Standard solution is stored in a light-proof box, ensuring quality.
– 5 cuvettes with 5 values 0nm, 420nm, 466nm, 525nm, 610nm.
CERTIFICATE OF ANALYSIS (COA) AND MSDS INCLUDED
– High quality HANNA standard solutions are manufactured in a state-of-the-art facility using pure chemicals in a temperature controlled environment
– Supplied with COA in each box.
– MSDS in file format.
SIMPLE AND FAST TO USE
1. Remove the cuvette from the opaque box.
2. Use the handle on the black lid, without opening the lid, use the HI731318 cuvette cleaning cloth to wipe the glass part. Cuvettes must not have any fingerprints or scratches, or stains.
3. See the CAL Check/Calibration section in the HI83300 meter's manual. Should be done at temperatures between 18 and 25 degreesoC.
Use until expiration date.
CONVENIENT STORAGE
– Store standard cuvettes upright in the original container, at temperatures from 5 to 30oC, avoid direct sunlight, do not freeze.
– ONLY for use with HANNA INSTRUMENTS devices.
- ARE NOT Open the lid of the cuvettes, dilute or add any chemicals to the solution.
– DO NOT extract or transfer the solution through another tool.