Force and mechanical measuring equipment
– 2 cuvettes already containing chlorine solution (residual and total) at 0 and 1 mg/L
– Supplied with a light-proof box, ensuring standard quality.
– Only used to check the accuracy and calibration of HANNA machines
– For use with HANNA HI96724 Photometer only
– Use until expiration date (>1 year).
– Supplied with COA and MSDS.
LIGHT-SHARDED CONTAINER BOX, AVOID COLLISION
– Standard solution is stored in a light-proof box, ensuring quality.
+ 1 cuvette A (white solution) at 0 mg/L
+ 1 cuvette B (pink solution) at 1 mg/L for residual chlorine and total chlorine
CERTIFICATE OF ANALYSIS (COA) AND MSDS INCLUDED
– High quality HANNA standard solutions are manufactured in a state-of-the-art facility using pure chemicals in a temperature controlled environment
– Supplied with COA in each box.
SIMPLE AND FAST TO USE
1. Take the 2 cuvettes out of the opaque box.
2. Use the handle on the black lid, without opening the lid, use the HI731318 cuvette cleaning cloth to wipe the glass part. Cuvettes must not have any fingerprints or scratches, or stains.
3. See CAL Check/Calibration section in the HI96724 meter manual. Should be done at temperatures between 18 and 25 degreesoC.
Use until expiration date
CONVENIENT STORAGE
– Store standard cuvettes upright in the original container, at temperatures from 5 to 30oC, avoid direct sunlight, do not freeze.
– ONLY for use with HANNA INSTRUMENTS devices.
- ARE NOT Open the lid of the cuvettes, dilute or add any chemicals to the solution.
– DO NOT extract or transfer the solution through another tool.