Force and mechanical measuring equipment
– 2 cuvettes containing standard chlorine solution (0 and 200 ppb)
– Pre-packaged, convenient.
– Used to check the accuracy of Checker HI761.
– Use until expiration date (>1 year).
– Supplied with COA and MSDS.
CONVENIENT PACKAGING
Comes pre-filled in HI761 checker cuvette
Just take it out and put it into the meter without needing to extract another tool.
Comes with Certificate of Analysis - Certificate of Analysis
High quality HANNA standard solutions are manufactured in a state-of-the-art facility using pure chemicals in a temperature controlled environment and are supplied with a COA in each box.
Reuse many times until expired
The expiration date is clearly printed on the container and can be used multiple times until expired.
VERY SIMPLE WAY TO CHECK THE MACHINE
Take 2 cuvettes out of the box. Use the handle on the black lid, without opening the lid, use the HI731318 cuvette cleaning cloth to wipe the glass part. Cuvettes must not have fingerprints, dirt or scratches.
1. Zero the machine with standard A cuvettes
2. Place standard cuvette B into the machine (C2)
3. Press the button and read the results on the screen. Compare the corresponding value on the accompanying Certificate of Analysis.
SIMPLE STORAGE
– Store the standard cuvette upright in the original container, at room temperature from 5 to 30 degreesoC, avoid direct sunlight, do not freeze.
– ONLY for use with HANNA INSTRUMENTS devices.
– DO NOT dilute or add any chemicals to the solution.
– DO NOT open the lid of the cuvettes.
– DO NOT extract or transfer the solution through another tool.
– DO NOT use expired products.