NIST Conductivity Standard Solution 1413µS/cm
Actual EC value table according to temperature is printed on the bottle body.
Long shelf life up to 4 years
Supplied with Certificate of Analysis (COA) for each bottle
H6031 is a high quality 1413µS/cm conductivity calibration solution standardized with a conductivity meter that has been calibrated with potassium chloride NIST SRM 2202 (Standard Reference Material) in deionized water for analytical use in accordance with ISO 3696 / BS3978.
HI6031 |
|
Value @25oC |
1413 µS/cm |
Specifications |
– 500mL black plastic bottle: Prevents any oxidation from UV light that could alter the buffer value |
Using |
– Pour enough calibration solution into a clean plastic cup so that the electrode, sensor and EC junction are completely immersed in the standard solution. – Calibrate the electrode according to the electrode or meter's instruction manual. |
Methods of preservation |
– After use, cover and store at room temperature, away from direct sunlight – Do not dilute or add any chemicals to the solution. – Always use fresh solution for each calibration – ABSOLUTELY do not pour used solution back into the bottle. |
Used for measuring machines |
Any conductivity meter can be calibrated at the point 1413 µS/cm |