– 2 cuvettes already contain test solution inside, no need to extract or mix anything else
– Only used to check the accuracy of the HI700 LR Ammonium Checker
– Supplied with COA and MSDS
*Checker does not have a calibration function, so the solution is only to check whether the machine is operating correctly or not
HOW TO CHECK CHECKER QUICKLY AND EASILY
Take 2 cuvettes out of the box. Use the handle on the black lid, without opening the lid, use the HI731318 cuvette cleaning cloth to wipe the glass part. Cuvettes must not have fingerprints, dirt or scratches.
1. Zero the machine with standard A cuvettes.
2. Place standard cuvette B into the instrument (C2).
3. Press the button and read the results on the screen. Compare the corresponding value on the accompanying Certificate of Analysis.
=> If the displayed results do not match the standard value, contact HANNA Technical (028.39260.457).
=> Checker does not have a calibration function, the machine only reads the results to see if they are still accurate or not.
SIMPLE STORAGE
– Store the standard cuvette upright in the original container, at room temperature from 5 to 30 degreesoC, avoid direct sunlight, do not freeze.
– ONLY for use with HANNA INSTRUMENTS devices
– DO NOT dilute or add any chemicals to the solution.
– DO NOT open the lid of the cuvettes.
– DO NOT extract or transfer the solution through another tool.
– DO NOT use expired products.