ORP Electrode Treatment (Reduction) Solution, 500mL HI7091L

Producer:
SKU: HANNA-HI7091L Category:

– Solution used to treat (reduce) the ORP electrode

– Expiry date is at least 1.5 years from the date of unopened packaging.

– Expiry date 1 month from date of opening

ORP (Oxidation Reduction Potential) electrodes are used to measure chemical reactions in which one species is being oxidized and another species is being reduced. The loss of electrons by an element or compound is called an oxidation reaction while the gain of electrons by an element or compound is called a reduction reaction. For example, electroplating companies have hexavalent chromium (Cr6+) in their wastewater must reduce it to the trivalent chromium form (Cr3+) safer. This reduction reaction occurs in the presence of a strong reducing agent such as sodium metabisulfite. The addition of the reducing agent must be measured with an ORP electrode to ensure that adequate reduction potential is present. Using the HI7091L reduction pretreatment solution prior to taking the measurement provides a faster response time in a reducing environment.

To use, soak the ORP electrode in the treatment solution for 15 minutes before use. As a general rule, if the mV value corresponding to the pH measurement is lower than the value in the table below the reducing solution (HI7091) is used. If the mV reading corresponding to the pH measurement is higher than the value in the table below, an oxidation treatment solution (HI7092) should be used.

 

pH 

mV 

pH

mV 

pH 

mV 

pH 

mV

pH 

mV 

0

990

3

800

6

640

9

460

12

280

1

920

4

740

7

580

10

400

13

220

2

860

5

680

8

520

11

340

14

160

 

 

HI7091L

Purpose

Used to treat ORP electrodes when the electrode needs to be calibrated when the corresponding mV value of pH is lower than the table above.

Specifications

1 bottle HI7091-1 500mL and 1 bottle HI7091-2 powder

Using

1. Pour the entire contents of the HI7091-2 powder bottle (small bottle) into the HI7091-1 solution bottle (large bottle).

2. Shake well 10 times to mix.

3. Use the mixed solution within 1 month.

4. Soak the ORP electrode in the treatment solution for 15 minutes before use.

Methods of preservation

– After use, cover and store at room temperature, away from direct sunlight

– Do not dilute or add any chemicals to the solution.

– Always use fresh solution for each calibration

– ABSOLUTELY do not pour used solution back into the bottle.

Used for measuring machines

Any ORP electrode 

Certificate of Analysis (COA)

Are not

MSDS

Have 

Guarantee

Are not

Request a quote
SEMIKI MEASUREMENT EQUIPMENT COMPANY LIMITED
Headquarters:
  • Floor 12 – Tower A2, Viettel Building, 285 Cach Mang Thang Tam, Ward 12, District 10, Ho Chi Minh City, Vietnam
  • Sales@semiki.com
  • +84 979761016
  • Tax code 0313928935
Warranty, maintenance and repair center:
  • 10th Floor, Halo Building, 19-19/2A Ho Van Hue, Ward 9, Phu Nhuan District, HCMC
Representative office in Hanoi:
  • 9th Floor, 3D Building, No. 3 Duy Tan Street, Dich Vong Hau Ward, Cau Giay District, Hanoi City
© 2024 Semiki inc. All rights reserved.

Ask SEMIKI for a quote

Complete the form below to receive a quote from SEMIKI.
You need to enter all information fields below.