Technical specifications of testo 526-2
Temperature – NTC | |
---|---|
Measuring range | -40 to +150 °C |
Accuracy | ±0.2 °C (-10 to +50 °C)*
±0.4 °C (Remaining range)* |
Resolution | 0.1°C |
*Accuracy information applies only to instrument without probes connected
Temperature – TC Type K (NiCr-Ni) | |
---|---|
Measuring range | -200 to +1370 °C |
Accuracy | ±0.4 °C (-100 to +200 °C)*
±1 °C (Remaining range)* |
Resolution | 0.1°C |
*Accuracy information applies only to instrument without probes connected
Differential pressure (internal sensor) – Piezoresistive | |
---|---|
Measuring range | 0 to 2000 hPa |
Accuracy | ±0.05 % of fsv |
Resolution | 0.1 hPa |
Static pressure | 2000 hPa (pressure) |
Overload | 3000hPa |
Zeroing | up to 50 hPa |
Differential pressure (external probe) – Piezoresistive | |
---|---|
Measuring range | 0 to 2000 hPa |
Accuracy | ±0.1 % measured value* |
Resolution | 0.10 Pa (0638 1347)
0.01 hPa (0638 1547) 0.001 hPa (0638 1447) 0.1 hPa (0638 1847 / 0638 1647) |
*Accuracy information applies only to instrument without probes connected
Differential pressure (external probe) – Ceramic | |
---|---|
Measuring range | -1 to 400 bar |
Accuracy | ±0.2 % of fsv* |
Resolution | 0.01 bar |
*Accuracy information applies only to instrument without probes connected
General specifications | |
---|---|
Size | 219x68x50mm |
Operating temperature | 0 to +50 °C |
Shell material | ABS |
Measurement cycle | 0.04 seconds |
Battery type | 9V (6LR61) |
Battery life | Continuous operation with internal pressure sensor: 30 h; With rech. battery: 10 h; With carbon battery: 18 h |
Connect | Hose: inner Ø 4 mm outer Ø 6 mm |
Screen type | LCD (Liquid Crystal Display) |
Display function | with symbols, 7 segment display and point matrix |
Display updates | 2x per second, in fast measurement 4x per second |
Power supply | Battery/Rechargeable battery, 12 V power supply |
Interface | RS232 |
Memory | 100 kB; 25,000 measurement values |
Storage temperatures | -20 to +70 °C |
Weight | 300g |