TOPCON IM-600/IM-600M Digital luminance meter

Producer:

TOPCON IM-600/IM-600M illumination meter meets JIS type AA standards. It has high reliability and accuracy. The IM-600 is an easy-to-use illuminance meter capable of measuring over a wide range. This product is ideal for evaluating next-generation lighting systems, and there is another IM-600M version available. This TOPCON IM-600/IM-600M product is useful for small spaces and equipment installations.

Features TOPCON IM-600/IM-600M

■ Wide range measurement (0.005 – 999,000 lx)
■ Display unit and light sensing unit are Separable with optional cable(IM-600)
■ Multiple measurement function
■ Response time, screen flicker can be measured with oscilloscope
■ Zero calibration function without cap can be installed equipment

Principal uses

■ Development and design department for lighting equipment maker.
■ Lighting system performance evaluation for building constructor.
■ Management of light source for plant factories.
■ Maintenance and management of light in factory, school, office, etc.
■ As a measurement sensor of production line or a variety equipments.

SpecificationsTOPCON IM-600/IM-600M

Item Specification
JIS Class Conforms to the general AA class illuminance meter
(JIS C 1609-1-2006)
Photo detector Silicon photo diode
Measurement
Range
0.005 – 999,000 lx
Cosine Correction Characteristics 【IM-600 】
10°: Within ±1%
30°: Within ±1%
50°: Within ±5%
60°: Within ±5%
80°: Within ±10%
【 IM-600M 】
10°: Within ±1%
30°: Within ±2%
50°: Within ±6%
60°: Within ±7%
80°: Within ±25%
Relative Spectral Response 6% or less (*1)
Linearity ±2% of rdg. ±1 digit (Auto Range)
Measurement
Mode
Auto / Manual (6 ranges)
Measurement
Function
Standard measurement:
Illuminance(lx), Integrated illuminance(lx・h), Integration time(h), Luminous intensity(cd)
Deviation measurement:
Illuminance difference(lx), Illuminance ratio(%)
Calibration
Function
CCF (Color Correction Factor) setting
Analog Output 0 – 3V
Interface USB
Power Source Size AA dry-cell battery×2 or Dedicated adapter (Option)
Use conditions Temperature : -10 – +40℃,
Humidity : 85% RH or less (No condensation)
Outer dimensions
(W×D×H)
188×65×34mm, Sensor Unit (IM-600M): 16×21mm
Weight IM-600: 200g, IM-600M: 230g *Non-dry cell battery

Accessories

■ AC Adapter: ZV-42EX(220V), ZV-42US(110V)
■ Extension cable: 2m, 5m, 10m, 20m, 30m

 

Maintenance and calibration : 

► We also have calibration certificate attached ( optional )

► Please re-calibrate after use for a certain period through our purchase store. Local Vietnam : SEMIKI CORP

DISTRIBUTED BY:

Semiki instrumentation Co., Ltd

Email: sales@semiki.com

Office tel: +84 28 2253 3522

~DOING OUR BEST FOR YOU~

 

Request a quote
SEMIKI MEASUREMENT EQUIPMENT COMPANY LIMITED
Headquarters:
  • Floor 12 – Tower A2, Viettel Building, 285 Cach Mang Thang Tam, Ward 12, District 10, Ho Chi Minh City, Vietnam
  • Sales@semiki.com
  • +84 979761016
  • Tax code 0313928935
Warranty, maintenance and repair center:
  • 10th Floor, Halo Building, 19-19/2A Ho Van Hue, Ward 9, Phu Nhuan District, HCMC
Representative office in Hanoi:
  • 9th Floor, 3D Building, No. 3 Duy Tan Street, Dich Vong Hau Ward, Cau Giay District, Hanoi City
© 2024 Semiki inc. All rights reserved.

Ask SEMIKI for a quote

Complete the form below to receive a quote from SEMIKI.
You need to enter all information fields below.