Force and mechanical measuring equipment
The WiFi-equipped testo Saveris 2T3 data logger allows you to monitor the temperature in processes, in refrigerators, work rooms or storage rooms. The measured values can be transmitted and stored online to the Testo Cloud via WiFi. You can access these measurement data at any time using your smartphone or computer with an internet connection.
The testo Saveris 2T3 WiFi data logger is the easy, flexible and reliable solution for monitoring temperatures in your processes, warehouses, refrigerators or workrooms. Configuring the system is simpler than ever, all via a web browser.
Provide cover:
Note: To operate the testo Saveris 2-T3 data logger, you need to connect at least one optional probe.
A temperature monitoring system consists of at least one testo Saveris 2 data logger and a Testo Cloud account. In addition, the testo Saveris 2T3 can connect up to two thermocouple temperature probes (TC Type K, T, J). All temperature values are transmitted and stored online in the Testo Cloud via WLAN. The testo Saveris 2T3 has a wide measuring range from -200 to +1350 °C thanks to optionally connectable thermocouple probes. You will receive immediate alerts via email or SMS (optional order) when the measured values exceed the thresholds.
The display shows the current measurement values, limit value violations and remaining battery life. The standard batteries (AA) have a 12-month life and can be replaced by the user at any time. The testo Saveris 2T3 is equipped with an internal memory for up to 20,000 measurement values (for 2 channels) and an IP54 housing.
The Testo Cloud is an essential component for setting up and operating your testo Saveris 2 system. Here you can make all the settings for the testo Saveris 2 data logger, set limit values and analyse the measurement data. You can access it anywhere and at any time using your smartphone, tablet or PC with an Internet connection. You must first register at www.saveris.net to gain access to the Testo Cloud.
Depending on your individual needs, you can choose between the following two Testo Cloud packages:
Basic (free) | Advanced | |
Measuring cycle | 15 min. … 24 h | 1 min. … 24 h |
Communication cycle | 15 min. … 24 h | 1 min. … 24 h |
Data storage | Max. 3 months | Max. 2 years |
Reports | Manual (.pdf/.csv) | Manual (.pdf/.csv) Automatic (.pdf/.csv) |
Number of users per account | 1 | 10 |
Number of WiFi data loggers per account |
Unlimited | Unlimited |
Alarm options | Upper/lower alarm limits |
|
System notifications |
|
|
Email alarm | Yes | Yes |
SMS alarm | No |
|
Temperature – TC Type K (NiCr-Ni) | |
---|---|
Measuring range | -195 to +1350 °C |
Accuracy | ±(0.5 + 0.5 % measured value) °C |
Resolution | 0.1°C |
Temperature – TC Type T (Cu-CuNi) | |
---|---|
Measuring range | -200 to +400 °C |
Accuracy | ±(0.5 + 0.5 % measured value) °C |
Resolution | 0.1°C |
Temperature – TC Type J (Fe-CuNi) | |
---|---|
Measuring range | -100 to +750 °C |
Accuracy | ±(0.5 + 0.5 % measured value) °C |
Resolution | 0.1°C |
General specifications | |
---|---|
Weight | 240 |
Size | 95 x 75 x 30.5mm |
Operating temperature | -30 to +50 °C |
Protection level | IP54 |
Measurement cycle | Depends on Testo Cloud package; Basic: 15 min … 24 hrs / Advanced: 1 min … 24 hrs |
Door contact | – |
Communication cycle | Depends on Testo Cloud package; Basic: 15 min … 24 hrs / Advanced: 1 min … 24 hrs |
WLAN connection | Signal transmission: wireless, frequency band: 2.4 GHz, supported wireless LAN standards: IEEE 802.11 b/g/n and IEEE 802.1X; Possible encryption methods: without encryption, WEP, WPA, WPA2, WPA2 Enterprise – The data loggers communicate via the standard protocol MQTT and are SNTP time synchronization-capable. |
Connection channel | External temperature probe |
Power supply | 4 x AA AlMn batteries; power supply optional; for temperatures below -10 °C, please use Energizer 0515 0572 batteries |
Memory | 10,000 measurement values/channel |
Battery life | 12 months (typical value, depending on the wireless LAN infrastructure) at +25 °C, 15-minute measuring cycle and standard communication cycle at -30 °C, 15-minute measuring cycle and standard communication cycle with Energizer batteries 0515 0572 |
Storage temperatures | -40 to +70 °C (without battery) |