Semiki logo
HOTLINE: +84 979761016
0

Cart

No products in the cart.

Continue shopping
Home / Shop / Force and mechanical measuring equipment / Gas meter / Toxic gas detector / Exhaust gas meter – testo 330-2LL
Product portfolio

Exhaust gas meter – testo 330-2LL

Producer:
SKU: TESTO-0632-3307 Category:
  • O sensor2 and high quality CO, long life (up to six years); Other sensors (like NO sensor) optional to order
  • 4-year warranty for testo 330-2LL exhaust gas analyzer
  • No need to unplug the sensor from the chimney to perform Zero
  • CO measurements up to 30,000 ppm

Discontinued, Replaced by TESTO 300

The TÜV tested testo 330-2LL exhaust gas meter ensures that you can measure for a long time! Modern, easy-to-use analysis tools with fast, accurate, reliable measurement results.

  • 4 year warranty for testo 330-2LL
  • Save more with long-life sensors (LL stands for Long Life). Because of the O sensor2 and CO does not need to be replaced too often (usually every 6 years), helping you save maximum costs
  • Many other types of sensors can be ordered optionally (such as: NO, CO sensorslow , NOlow )
  • Innovative design makes changing sensors quick and easy.

testo 330-2LL exhaust gas meter – Application for a wide range of measurements

  • Measurements of exhaust gas parameters (CO, O2 , temperature) at the burner
  • Draft measurements at the heating system
  • Measure pressure at burners (mold pressure, gas flow pressure)
  • Measure ambient CO with an optional sensor to order
  • Detect gas leaks and check gas pipes with optional sensors
  • Integrated differential pressure measurement

What is inside the testo 330-2LL exhaust gas meter?

  • Includes O sensors2 , CO (long life) and NO
  • Scientific sensor slot design makes it easy to change
  • High-resolution graphical color display, presenting the measurement process as a graph
  • Multiple measurement menus are stored in the device (just open and measure)
  • Integrated sensor monitoring feature; Light coding system (red, blue, yellow) helps you understand sensor status at a glance
  • User can change the sensor; Easy sensor filter
  • No need to retract the probe from the chimney to Zero (additional feature that testo 330-1LL does not have)
  • Automatic CO dilution from 8,000 ppm to at least 30,000 ppm (another additional feature that the testo 330-1LL does not have)
  • Data logging function for long-term measurements; Replaceable memory for up to 500,000 measurement values
  • The integrated magnet on the back helps position the testo 330 on metal surfaces
  • Integrated dew trap
  • Bluetooth connectivity for testo BLUETOOTH® printers (optional to order)
  • Device warranty is up to 4 years
  • TÜV-tested according to 1st BImSchV EN 50.379

Specifications

Differential pressure – Piezoresistive
Measuring range ±10,000 Pa
Accuracy ±0.3 Pa (0 to 9.99 Pa) ±1 digit

±3 % measured value (10 to 10,000 Pa) ±1 digit

O₂ gas
Measuring range 0 to 21 Vol.%
Accuracy ±0.2 Vol.%
Resolution 0.1 Vol.%
Response time t₉₀ < 20 seconds
CO gas (H₂ compensation)
Measuring range 0 to 8,000 ppm
Accuracy ±10 ppm or ±10 % measurement value (0 to 200 ppm)

±20 ppm or ±5 % measured value (201 to 2,000 ppm)

±10 % measured value (2,001 to 8,000 ppm)

Resolution 1 ppm
Response time t₉₀ < 60 seconds
CO gaslow
Measuring range 0 to 500 ppm
Accuracy ±2 ppm (0 to 39.9 ppm)

±5 % measured value (40 to 500 ppm)

Resolution 0.1 ppm
Response time t₉₀ < 40 seconds
CO determination (H₂ compensation), automatic dilution
Measuring range 0 to 30,000 ppm
Accuracy ±100 ppm (0 to 1000 ppm)

±10 % measured value (1,001 to 30,000 ppm)

Resolution 1 ppm
NO gas
Measuring range 0 to 3,000 ppm
Accuracy ±5 ppm (0 to 100 ppm)

±5 % measured value (101 to 2,000 ppm)

±10 % measured value (2,001 to 3,000 ppm)

Resolution 1 ppm
Response time t₉₀ < 30 seconds
Furnace gas NOlow
Measuring range 0 to 300 ppm
Accuracy ±2 ppm (0 to 39.9 ppm)

±5 % measured value (40 to 300 ppm)

Resolution 0.1 ppm
Response time t₉₀ < 30 seconds
Emissions
Measuring range -9.99 to +40 hPa
Accuracy ±0.02 hPa or ±5 % measured value (-0.50 to + 0.60 hPa)

±0.03 hPa (+0.61 to +3.00 hPa)

±1.5 % measured value (+3.01 to +40.00 hPa)

Resolution 0.01 hPa
Temperature
Measuring range -40 to +1200 °C
Accuracy ±0.5 °C (0 to +100.0 °C)

±0.5 % measurement value (Remaining measurement range)

Resolution 0.1 °C (-40 to +999.9 °C)

1 °C (> +1,000 °C)

Furnace efficiency, Eta (calculated)
Measuring range 0 to 120 %
Resolution 0,1 %
Exhaust gas loss (estimated)
Measuring range 0 to 99.9 %
Resolution 0,1 %
CO gas2 of the furnace (calculated from O₂)
Measuring range 0 to CO₂ max (Display range)
Accuracy ±0.2 Vol.%
Resolution 0.1 Vol.%
Response time t₉₀ < 40 seconds
Velocity/flow
Measuring range 0.15 to 3 m/s
Resolution 0.1 m/s
CO gas (without H₂ compensation)
Measuring range 0 to 4,000 ppm
Accuracy ±20 ppm (0 to 400 ppm)

±5 % measured value (401 to 2,000 ppm)

±10 % measured value (2,001 to 4,000 ppm)

Resolution 1 ppm
Response time t₉₀ < 60 seconds
Ambient CO
Measuring range 0 to 500 ppm
Accuracy ±5 ppm (0 to 100 ppm)

±5 % measured value (> 100 ppm)

Resolution 1 ppm
Response time About 35 seconds

With CO probe

Ambient CO₂
Measuring range 0 to 1 Vol.%

0 to 10,000 ppm

Accuracy ±50 ppm or ±2 % measurement value (0 to 5,000 ppm)

±100 ppm or ±3 % measured value (5,001 to 10,000 ppm)

Response time About 35 seconds

With ambient CO probe

Gas leak detection for flammable gases (via gas leak detector)
Measuring range 0 to 10,000 ppm CH₄ / C₃H₈; Display range
Accuracy LED warning signal, buzzer signal
Response time t₉ ₀ < 2 s

With gas leak detector

Temperature (via fine pressure probe)
Measuring range -40 to +1200 °C maximum (Depends on probe)
Accuracy ±0.5 °C (-40 to 100 °C)

±0.5 % measured value (remaining measuring range) + probe accuracy

Resolution 0.1 °C
General specifications
Size 270 x 90 x 65 mm
Operating temperature -5 to +45 °C
Guarantee Main unit: 12 months, sensor and probe (Optional)
Display size 240 x 320 pixels
Display Color screen
Power supply Rechargeable battery 3.7 V / 2.6 Ah; Circuit 6 V / 1.2 A
Maximum memory 500,000 measurement values
Storage temperatures -20 to + 50 °C
Weight 600 g (without rechargeable battery)

>>> Youtube: Here

>>> Fanpage: Here

>>> News: Here

Request a quote
SEMIKI MEASUREMENT EQUIPMENT COMPANY LIMITED
Headquarters:
  • Floor 12 – Tower A2, Viettel Building, 285 Cach Mang Thang Tam, Ward 12, District 10, Ho Chi Minh City, Vietnam
  • Sales@semiki.com
  • +84 979761016
  • Tax code 0313928935
Warranty, maintenance and repair center:
  • 10th Floor, Halo Building, 19-19/2A Ho Van Hue, Ward 9, Phu Nhuan District, HCMC
Representative office in Hanoi:
  • 9th Floor, 3D Building, No. 3 Duy Tan Street, Dich Vong Hau Ward, Cau Giay District, Hanoi City
© 2024 Semiki inc. All rights reserved.

Ask SEMIKI for a quote

Complete the form below to receive a quote from SEMIKI.
You need to enter all information fields below.