General specifications | |
---|---|
Weight | 960 g |
Size | 283 x 103 x 65 mm |
Operating temperature | -5 to +50°C |
Shell material | TPEPC |
Protection level | IP40 |
Connectable probe | 1x exhaust detector;
1x temperature probe; 1x differential pressure |
Product color | Black |
Battery life | > 6h (pump on, display light off, at 20 °C) |
Display screen | Graphic display |
Screen size | 160 x 240 pixels |
Display function | Graphic display |
Power supply | Battery pack 3.7 V / 2.4 Ah, Power supply 6.3 V / 2 A |
User-selectable fuel | 10 user-customizable fuel types |
Pump flow | 0.6 liters/minute (regulatory) |
Tube length | maximum 7.8 m (corresponding to two probe tube extensions) |
Max. pos. press./flue gas | 50 mbar |
Max. neg. neg. press./flue gas | -200 mbar |
Maximum memory | 100 folders |
Memory per folder | Maximum 10 sites |
Storage per site | Maximum 200 records |
Miscellaneous | The max. number of logs is determined by the number of folders or sites |
Display | Bluetooth®; USB; IR/IRDA interface; gas outlet; Mains connection;
probe input/ probe inputs; Differential Pressure |
Storage temperatures | -20 to +50 °C |
Differential pressure – Piezoresistive | |
---|---|
Measuring range | -200 to 200 hPa |
Accuracy | ±0.5 hPa (-49.9 to 49.9 hPa)
±1.5 % measured value (Remaining measuring range) |
Resolution | 0.1 hPa |
Absolute pressure | |
---|---|
Measuring range | +600 to +1150 hPa |
Accuracy | ±10 hPa |
Resolution | 1 hPa |
O₂ gas | |
---|---|
Measuring range | 0 to 25 Vol.% |
Accuracy | ±0.2 Vol.% |
Resolution | 0.01 Vol.% |
Response time t₉₀ | <20 seconds |
CO gas (with H₂ compensation) | |
---|---|
Measuring range | 0 to 10,000 ppm |
Accuracy | ±10 ppm or ±10 % measurement value (0 to 200 ppm)
±20 ppm or ±5 % measured value (201 to 2000 ppm) ±10 % measured value (2001 to 10,000 ppm) |
Resolution | 1 ppm |
Response time t₉₀ | <40 seconds |
COlow (with H₂ compensation) | |
---|---|
Measuring range | 0 to 500 ppm |
Accuracy | ±2 ppm (0 to 39.9 ppm)
±5 % measuring value (Remaining measuring range) at ambient temperature of 20 °C. Additional temperature coefficient 0.25 % measured value/K. |
Resolution | 0.1 ppm |
Response time t₉₀ | <40 seconds |
NO emissions | |
---|---|
Measuring range | 0 to 4000 ppm |
Accuracy | ±5 ppm (0 to 99 ppm)
±5 % measured value (100 to 1999 ppm) ±10 % measured value (2000 to 4000 ppm) |
Resolution | 1 ppm |
Response time t₉₀ | <30 seconds |
Nolow emissions | |
---|---|
Measuring range | 0 to 300 ppm |
Accuracy | ±2 ppm (0 to 39.9 ppm)
±5 % measuring value (Remaining measuring range) |
Resolution | 0.1 ppm |
Response time t₉₀ | <30 seconds |
NO₂ emissions | |
---|---|
Measuring range | 0 to 500 ppm |
Accuracy | ±10 ppm (0 to 199 ppm)
±5 % measuring value (Remaining measuring range) |
Resolution | 0.1 ppm |
Response time t₉₀ | <40 seconds |
To avoid absorption, a maximum measurement duration of 2 hours should not be exceeded.
SO₂ emissions | |
---|---|
Measuring range | 0 to 5000 ppm |
Accuracy | ±10 ppm (0 to 99 ppm)
±10 % measurement value (Remaining measurement range) |
Resolution | 1 ppm |
Response time t₉₀ | <40 seconds |
To avoid absorption, a maximum measurement duration of 2 hours should not be exceeded.
Differential Pressure Draft – Piezoresistive | |
---|---|
Measuring range | -40 to +40 hPa |
Accuracy | ±0.03 hPa (-2.99 to +2.99 hPa)
±1.5 % measured value (Remaining measuring range) |
Resolution | 0.01 hPa |
Temperature | |
---|---|
Measuring range | -40 to +1200 °C |
Accuracy | ±0.5 °C (0 to +99 °C)
±0.5 % measurement value (Remaining measurement range) |
Resolution | 0.1 °C |
Efficiency level of emissions, Eta (calculated) | |
---|---|
Measuring range | 0 to 120% |
Resolution | 0,1% |
Exhaust gas escape (calculation) | |
---|---|
Measuring range | 0 to 99.9% |
Resolution | 0,1% |
Dew point (calculated) | |
---|---|
Measuring range | 0 to 99.9 °Ctd |
Resolution | 0.1 °Ctd |
CO₂ calculation (calculated from O₂) | |
---|---|
Measuring range | 0 to CO₂ max |
Accuracy | ±0.2 Vol.% |
Resolution | 0.1 Vol.% |
Response time t₉₀ | <40 seconds |