The HI931101 is a durable, waterproof, portable sodium meter. The HI931101 includes many features such as automatic calibration, automatic temperature compensation, auto ranging, and on-screen instructions. There are four scales for sodium and simply press the MODE button to measure the activity of sodium ions.
The device does not come with electrodes and calibration solution (purchased separately).
+ FC300B: salt electrode + HI7662: temperature probe
+ HI7080L + HI7086L + HI7087L Standard Solution
+ HI7090L Ion Adjustment ISA Solution for High Salinity
+ ISA Solution Ion Adjustment for Low Salinity HI4016-00
+ HI4016-46 Electrode Cleaning Solution
+ HI4016-45 Sodium Electrode Storage Solution
+ Solution sodium electrode body HI7079
The HI931101 is a rugged, portable sodium meter designed to provide laboratory accuracy and precision under harsh conditions. The HI931101 is ideal for a variety of applications in the laboratory, food and beverage manufacturing, and water quality analysis. The HI931101 offers a variety of features including 5 different measurement ranges, automatic calibration to 1 or 2 points, temperature compensation, and on-screen instructions.
The HI931101 uses a BNC connector for connecting a sodium ISE electrode and a separate RCA connector for a temperature probe. The temperature probe allows for high measurement accuracy of ±0.2oC to accurately measure temperature.
Outstanding Features
One or two point calibration – Automatic calibration can be performed at 1 or 2 points with a choice of 3 buffers (0.23 g/L, 2.3 g/L and 23.0 g/L). For high accuracy, the calibration solution should be selected at the temperature closest to the sample temperature.
On-screen instructions – Clear and easy-to-use on-screen instructions guide you through all the steps of performing calibration and measurement.
Sodium ion selective electrode – The FC300B sodium ion selective electrode consists of a glass body and a sodium sensitive membrane. The sodium sensitive membrane should be conditioned before calibration. Conditioning the electrode is a two-step process in which the glass is acid etched and then hydrated. Etching involves removing a thin layer of the sensitive glass from the sodium membrane, exposing a new sodium-containing surface. Hydration after etching will optimize the exchange of sodium ions between the membrane and the solution.
Fixed Temperature Compensation – HI931101 compensates for fixed temperature of sodium measurements at 25°C.
The device does not come with electrodes and calibration solution (purchased separately).
SPECIFICATIONS
The scale |
0.00 to 3.00 pNa; 15.0 to 150.0 mg/L (ppm) Na; 0.150 to 1.500 g/L Na; 1.50 to 15.00 g/L Na; 15.0 to 60.0 g/L Na |
-20 to 120oC (-4.0 to 248.0°F) |
|
Resolution |
0.01 pNa; 0.1 mg/L (ppm) Na; 0.001 g/L Na; 0.01 g/L Na; 0.1 g/L Na; 1 g/L Na |
0.1oC (0.1°F) |
|
Accuracy (@20oC) |
±0.05 pNa; ±5% measured value (Na) |
± 0.5oC (excluding probe error) |
|
EMC deviation |
± 2% FS (pNa and Na) / ± 0.5oC |
Calibration |
automatic, 1 or 2 points at 2.3 g/l (HI7080L), 0.23 g/l (HI7087L), 23.0 g/l (HI7086L) |
Heat compensation |
default at 25oC (77°F) |
NaCl measuring electrode |
FC300B, 1m cable (purchased separately) |
Temperature probe |
HI7662, 1 m cable (sold separately) |
Input impedance |
1012 Ohm |
Electricity supply |
4 1.5V batteries |
Environment |
0 – 50oC; RH max 100% |
Size |
196 x 80 x 60 mm |
Mass |
500 g |
Guarantee |
12 months for machine |
Offer includes |