Force and mechanical measuring equipment
– Membrane dissolved oxygen measurement
– Suitable for measuring rivers, lakes, and still water
– Manual calibration
– Polarographic electrode
– Replace membrane cap and filling solution once a month
– Comes with full accessories and shock-absorbing suitcase for field trips
POLAROGRAPHIC MEASURING SYSTEM
The meter and probe use Polarographic sensor technology based on the Ross and Clark polarographic measurement method. The probe consists of a platinum cathode and silver anode in an electrolyte solution held in place on the surface of the electrode by a polymer membrane. Oxygen diffuses through the membrane and establishes a current proportional to the concentration of dissolved oxygen.
MANUAL CALIBRATION
The meter has manual calibration feature at 100% in air
AUTOMATIC TEMPERATURE COMPENSATION
All values are automatically compensated for temperature with high precision, built-in temperature sensor behind the stainless steel case.
REPLACE MEMBRANE CAP AND FILLING SOLUTION PERIODICALLY
– The PTFE membrane cap uses a screw cap design that can be quickly replaced by filling with HI7041 electrolyte solution and screwing onto the DO probe.
– Periodically once a month.
– The machine comes with 2 membrane caps and 1 30mL filling solution
VIDEO
SPECIFICATIONS
DO scale |
0.0 to 19.9 mg/L (ppm) |
DO Resolution |
0.1 mg/L (ppm) |
DO Accuracy |
±1.5% FS (@ 25°C/77°F) |
Temperature scale |
-5.0 to 50.0°C (23.0 to 122.0°F) |
Temperature resolution |
0.1°C (1°F) |
Temperature accuracy |
±0.2°C (±1°F) (excluding probe error) (@ 25°C/77°F) |
Temperature compensation |
Automatic, 0 to 50°C (32 to 122°F) |
Calibration |
Manually, at one or two points (zero and slope) |
Electrode |
Polarographic type with integrated temperature sensor, DIN connector |
The battery |
3 1.5V batteries / about 1000 hours of use |
Environment |
0 to 50°C (32 to 122°F); RH max 100% |
Size |
196 x 80 x 60 mm |
Mass |
500 g |
Guarantee |
12 months for the machine and 6 months for the electrode (for products with intact tags and warranty card) |
Supply includes |
– Measuring machine HI9142 – Dissolved oxygen electrode HI76407/4 4m cable – 2 PTFE DO membrane caps HI76407A/P – Electrolytic solution HI7041S (30 ml) – Battery. - User manual. - Warranty. – Carrying case. |