Force and mechanical measuring equipment
- Measure Bromine, Chlorine, Cyanuric Acid, Iodine, Iron, pH and Turbidity of Clean Water
– All in one meter
- Measure each indicator, buy separate reagents for that indicator.
– Validation and Calibration with CAL Check™
– Countdown timer when measuring
– Convenient handheld meter suitable for on-site measurement.
IMPORTANCE
Turbidity, free and total chlorine, cyanuric acid, pH, iodine, bromine, and low-range iron are all important parameters that need to be tested to ensure water quality in drinking water, wastewater, and water used for swimming pools and spas.
Turbidity is one of the most important parameters used to determine the quality of drinking water. Turbidity in drinking water can be considered as a pathogenic factor. In natural waters, turbidity is measured to assess the general quality of water and its suitability in applications involving aquatic organisms.
Chlorine is the most commonly used water disinfectant in homes, hotels, and businesses. In swimming pools, spas, and similar applications, cyanuric acid helps stabilize and prevent chlorine from degrading, especially in sunlight. Regularly testing both cyanuric acid and pH levels helps minimize chlorine consumption.
Bromine and iodine are considered alternatives to chlorine for their disinfectant properties. One of the most common uses of iodine is in the poultry industry, while bromine is used in the sanitization of cooling towers, swimming pools, and spas.
Iron is naturally present in low concentrations in water, but it reaches high concentrations in wastewater. Iron concentrations in water need to be monitored because it becomes harmful above certain levels. In water, for example, iron can stain fabrics, damage kitchenware, promote the growth of certain bacteria, and change the taste of water. Iron is also an indicator of ongoing corrosion in cooling and heating water systems. Furthermore, iron is often monitored in wastewater to prevent contamination.
ALL IN 1 MULTI-FUNCTIONAL METER
– To measure turbidity, no reagents required.
– For Bromine, Chlorine, Cyanuric Acid, Iodine, Iron and pH, just buy the reagents for the indicators to be measured.
TURBIDITY CALIBRATION
HI93102 2-point turbidity calibration at 0 and 20 NTU (standard solution purchased separately)
GLP
Review last calibration information and date
LOCK SYSTEM ON CUVET CAP AND MACHINE
There is a marking on the measuring chamber of the machine and on the cuvette cover. Make sure the cuvettes are inserted into the measurement chamber in the same position each time you measure.
DATA LOG
Records up to 25 measurement values. Records can be easily recalled with the push of a button and a warning will appear if the memory is full. Logging can be enabled or disabled.
ERROR WARNING
SPECIFICATIONS
pH |
|
The scale |
5.9 to 8.5 pH |
Resolution |
0.1 pH |
Accuracy |
± 0.1 pH |
Reagents |
HI93710-01 |
Turbidity |
|
The scale |
0.00 to 50.0 NTU |
Resolution |
0.01 (0.00 to 9.99); 0.1 NTU (10.0 to 50.0) |
Accuracy |
±0.5 NTU or ±5% measurement result |
Calibration |
2 points, 0.00 and 20.0 FTU |
Calibration solution |
HI93102-0: 0 NTU HI93102-20: 20 NTU |
Residual chlorine and total chlorine |
|
The scale |
Residual Chlorine: 0.00 to 2.50 mg/L (ppm) Cl2 Total Chlorine: 0.00 to 3.50 mg/L (ppm) Cl2 |
Resolution |
0.01 mg/L |
Accuracy |
±0.03 mg/L ±3% measurement result |
Reagents |
Residual chlorine: HI93701-01 Total chlorine: HI93711-01 |
Cyanuric Acid |
|
The scale |
0 to 80 mg/L (ppm) |
Resolution |
1mg/L |
Accuracy |
±1 mg/L ±15% measurement result |
Reagents |
HI93722-01 |
Iodine |
|
The scale |
0.0 to 12.5 mg/L (ppm) I2 |
Resolution |
0.1 mg/L |
Accuracy |
±0.1 mg/L ±5% measurement result |
Reagents |
HI93718-01 |
Bromine |
|
The scale |
0.00 to 8.00 mg/L (ppm) Br2 |
Resolution |
0.01 mg/L |
Accuracy |
±0.08 mg/L ±3% measurement result |
Reagents |
HI93716-01 |
Low ladder iron |
|
The scale |
0.00 to 1.00 mg/L (ppm) Fe |
Resolution |
0.01 mg/L |
Accuracy |
±0.02 mg/L ±3% measurement result |
Method |
TPTZ |
Reagents |
HI93746-01 |
Measuring machine parameters |
|
Light source |
LED @525nm |
Record data |
Up to 25 copies |
Auto Shutdown |
after 10 minutes in measurement mode; after 1 hour in standard mode since last measurement |
Environment |
0 to 50°C (32 to 122°F); RH max 95% non-condensing |
The battery |
4 x 1.5V |
Size |
220 x 82 x 66 mm |
Mass |
510 g |
Guarantee |
12 months (for products with original stamps and warranty cards intact) |
Supply includes |
– HI93102 meter with battery – 2 cuvettes with lids – Warranty card (no warranty without warranty card) – Quality certificate of measuring machine - User manual – Paper box Turbidity calibration solution and reagents purchased separately |