Force and mechanical measuring equipment
– Modern handheld pH/ORP/ISE/Temperature meter.
– CAL Check feature alerts when the electrode needs cleaning or calibration
– Automatic data recording
– Automatic pH calibration up to 5 points
– Connect computer via USB
– The device comes with a pH electrode and pH standard solution.
– Sturdy, professional case
AUTOMATIC DATA RECORDING
Manually log data on demand for up to 200 samples (100 for pH and 100 for mV).
GLP
With the GLP button, the screen will display the recorded standard data including the standard value, the most recent date, month and year.
BNC PORT FOR CONNECTING ELECTRODES
– HI98191 meter is supplied with HI72911B pH/temperature electrode with titanium body and BNC connector with 1 m cable, HI7662 temperature probe
– The meter can be connected to any ORP electrode with a BNC port.
– The HI98191 can connect to 15 different ion selective sensors (ISEs) that are pre-programmed into the meter. By selecting the appropriate ISE electrode in the ISE setup, the appropriate ion valence and molecular weight are applied to display the ISE slope percentage. The user can then calibrate up to 5 points with a choice of 6 standard values. The meter allows selection of measurement units (ppm, ppt, g/L, ppb, mg/L, mg/ml, M, mol/L, mmol/L,% w/v, user) and a wide measuring range from 1.00 x 10-7 up to 9.99 x 1010.
LCD DISPLAY WITH BACKLIGHT
Supports pH measurement in low light conditions.
ERROR SCREEN
The display shows error warnings during calibration or measurement to alert the user.
AUTOMATIC CALIBRATION
Simply press the CAL button and immerse the electrode in the standard solution, the meter will automatically recognize the buffer and calibrate up to 5 points with available points: 1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45 and 5 custom buffers
CAL Check™ alerts when the electrode needs cleaning or recalibration
– After cleaning and calibration, the probe status is displayed as a percentage from 0-100% in 10% increments.
– To compare previous calibration data with the current standard, the meter will alert the user with a reminder display when the probe needs to be cleaned or is damaged, or the pH buffer may be contaminated.
BNC PORT FOR ELECTRODE CONNECTION
Helps connect quickly, simply and without twisting or tangling between wires.
USB CONNECTION
The HI98191 can transfer data directly to a PC using the USB cable and HI92000 software which is available as a free download from HANNA.
PROVIDE FULL SET OF ACCESSORIES
HI98191 comes with complete accessories at no additional cost.
– HI98191 meter with battery
– HI72911B pH/temperature electrode 1m cable
- DHI7662 temperature probe
– pH4.01 and pH7.01 calibration solution
– Convenient and sturdy suitcase for carrying the device
ISE ELECTRODE AND NECESSARY SOLUTIONS AND ACCESSORIES
https://hannavietnam.com/blog-detail/dien-cuc-ise-va-cac-dung-dich-phu-kien-can-thiet-122
SPECIFICATIONS
pH scale |
-2.0 to 20.0 pH -2.00 to 20.00 pH -2,000 to 20,000 pH |
pH resolution |
0.1 pH; 0.01 pH; 0.001 pH |
pH accuracy |
±0.1; ±0.01 pH; ±0.002 pH |
pH calibration |
up to 5 points with 7 available standards (1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45) and 5 custom buffers |
pH electrode |
HI72911B titanium body with internal temperature sensor, BNC connector, 1 m cable (included) |
ORP scale |
±2000 mV |
ORP resolution |
0.1 mV |
ORP Accuracy |
±0.2 mV |
ORP electrode |
BNC port (purchased separately) |
ISE scale |
from 1.00 E⁻⁷ to 9.99 E¹⁰ concentration |
ISE resolution |
3 digits 0.01; 0.1; 1; 10 concentration |
ISE Accuracy |
±0.5% value (I-valence), ±1% value (II-valence) |
ISE Calibration |
up to 5 points with 6 available standards (0.1, 1, 10, 100, 1000, 10000 ppm) |
ISE electrode |
BNC port (purchased separately) |
Temperature scale |
-20.0 to 120.0°C (-4.0 to 248.0°F) |
Temperature resolution |
0.1°C (0.1°F) |
Temperature accuracy |
±0.4°C (±0.8°F) (excluding probe error) |
Temperature compensation |
Manual or automatic from -20.0 to 120.0°C (-4.0 to 248.0°F) |
Slope Calibration |
From 80 to 110% |
Input |
10¹² Ohms |
Record on request |
300 samples (100 each for pH, mV and ISE) |
PC connection |
USB port with HI 92000 software and USB cable |
The battery |
4 1.5V AA batteries / approximately 200 hours of continuous use without backlight (50 hours with backlight) |
Automatically turns off |
Options: 5, 10, 30, 60 minutes or disabled |
Environment |
0 to 50°C (32 to 122°F); RH 100% (IP67) |
Size |
185 x 93 x 35.2 mm |
Mass |
400 g |
Guarantee |
12 months for the machine and 06 months for the included electrodes. (for products with intact tags and warranty card) |
Supply includes |
– HI98191 meter – HI72911B pH and temperature combination electrode – HI7662 temperature probe – HI7004M pH4.01 Standard Solution (230 mL) – HI7007M pH7.01 Standard Solution (230 mL) – 2 packets of HI700601 electrode cleaning solution – 4 AA 1.5V batteries (inside the machine) – 100mL plastic cup - USB cable - User manual – Quality certificate for machine and electrodes – Suitcase for carrying the device |