Force and mechanical measuring equipment
– Measure pH and Temperature simultaneously.
– ORP can be measured with a separately purchased ORP electrode.
– Automatic calibration at 2 points
– Automatic temperature compensation.
– HI8424 meter comes with HI1230B pH electrode, HI7662 temperature probe, and calibration solutions.
– Convenient storage box
AUTOMATIC CALIBRATION
At 1 or 2 points with 3 pre-programmed points (pH 4.01,7.01, 10.01)
HOLD BUTTON
The device has a HOLD button to help keep the pH value stable on the screen.
AUTOMATIC TEMPERATURE COMPENSATION
The HI8424 offers automatic temperature compensation from -20.0 to 120.0°C with the included HI7662 temperature probe.
The meter can also be compensated manually by entering the current sample temperature value into the meter when the HI7662 temperature probe is not present.
PH ELECTRODE WITH DURABLE PLASTIC BODY
Gel electrode without electrolyte filling, double junction.
The double junction design reduces internal reference cell interference that can cause precipitation of Ag/AgCl electrolytes leading to clogged junctions.
BNC CONNECTOR
The HI8424 has a BNC connector that is compatible with all electrodes that have a corresponding BNC connector.
CONVENIENT CARRYING BOX
HI8424 is supplied with HI1230B pH electrode, HI7662 temperature probe, pH 4.01 and pH 7.01 buffer solution sachets, batteries, instructions and rugged carrying case.
ORP ELECTRODE purchased separately when ORP measurement is required
SPECIFICATIONS
pH scale |
-2.00 to 16.00 pH |
pH resolution |
0.01 pH |
pH accuracy |
±0.01 pH |
pH calibration |
automatic, at 1 or 2 points (pH 4.01, 7.01, 10.01) |
ORP scale |
±699.9 mV; ±1999 mV |
ORP resolution |
0.1 mV; 1 mV |
ORP Accuracy |
±0.2 mV; ±1 mV |
Temperature scale |
-20.0 to 120.0°C / -4.0 to 248.0°F |
Temperature resolution |
0.1°C / 0.1°F |
Temperature accuracy |
±0.4°C; ±0.8°F |
Heat compensation |
automatically from -20.0 to 120.0ºC (-4.0 to 248.0ºF) or manually without connecting a temperature probe |
pH electrode |
HI1230B PEI body, BNC connector and 1 m cable (included) |
ORP electrode |
HI3230B PEI body, BNC connector and 1 m cable (sold separately) HI3131B glass body, BNC connector and 1 m cable (sold separately) |
Temperature probe |
HI7662 stainless steel with 1 m cable (included) |
Slope calibration |
From 75 to 110% / ±1 pH |
Input |
10¹² Ohms |
The battery |
1 9V battery/equivalent to 150 hours of continuous use, automatically turns off after 20 minutes of non-use |
Environment |
0 to 50°C (32 to 122°F); RH max 100% |
Size |
164 x 76 x 45 mm |
Mass |
180 g |
Guarantee |
12 months for the machine and 6 months for the electrode (For products with intact tags and warranty card) |
Supply includes |
– Measuring machineHI8424 – pH electrode HI1230B plastic body 1m cable – Temperature probe HI7662 1m cable – Machine standard solution package pH4.01 HI70004P – Machine standard solution package pH7.01 HI70007P – Solution package wash HI700601P electrode – The battery (in machine) - Warranty - User manual – Quality certificate for machine and probe – Plastic box containing the machine. *ORP electrode purchased separately |