Home / Shop / PCA330-2 Chlorine, pH, ORP and Temperature Analyzer and Control
Product portfolio

PCA330-2 Chlorine, pH, ORP and Temperature Analyzer and Control

Producer:
SKU: HANNA-PCA330-2 Category:

HANNA PCA330 series analyzes and controls chlorine concentration, pH, ORP and temperature continuously in-line and adjustable chlorine dosing, optional sampling time, alarm system, data connection via GSM network, easy-to-use interface, serial communication via RS485, recording output, 4-20mA dosing output, NEMA 4X housing

From drinking water and wastewater treatment to pool and spa sanitation, controlling chlorine, bromine or iodine levels is important for public health as well as ROI and beneficial for heating systems and industrial applications.

The PCA330 series controls residual chlorine or total chlorine from 0.00 to 5.00 mg/L

In the DPD colorimetric method, the indicator N, N-Diethyl-p-phenylene-diamine and a buffer are mixed with the sample. The residual chlorine oxidizes the DPD indicator reagent at pH 5.5 and 6.0 to form a red compound. The intensity of the color is proportional to the concentration of chlorine in the sample. The purpose of the buffer solution is to maintain the appropriate pH.

To measure total chlorine (free chlorine and combined chlorine), potassium iodide is added. A pH of 5.1 is required for this reaction. Thus, total chlorine measurement requires another buffer solution containing potassium iodide. Once the reactions are complete, the optical signal at 555 nm is compared with the signal obtained after passing the sample before the reagent was added. Finally, the chlorine concentration is calculated and displayed on the screen.

Indicator and buffer reagents are supplied with the meter. With a 5 minute sampling time, reagents only need to be replenished about once a month.

Since chlorine, bromine and iodine are closely related to pH, HANNA has designed the new PCA320 analyzer to monitor pH and temperature. degree and PCA330 for pH, ORP and temperature control .

The PCA320 and PCA330 use the HI1005 pH probe to continuously measure pH in the sample stream over a range of 0 to 14 pH. Sample temperature is measured over a range of 5 to 75°C. pH and temperature are displayed on the front panel, and the pH value is temperature compensated.

The PCA330 uses the HI2008 platinum ORP electrode to continuously measure the ORP value of the sample. A combination pH/temperature sensor and ORP sensor are integrated inside for direct in-stream measurement of the sample.

Measurement and control cycle

PCA has a control time cycle that can be set by the user depending on the size of the control system. Control, pumping and alarm processes can be performed based on this time cycle. The cycle can be set from 3 to 90 minutes.

Chlorine Control

Four user-adjustable chlorine level setpoints: one proportional dosing setpoint, two alarm setpoints and minimum pump level. The proportional dosing factor (1/Δ) is optional with Δ between 0.1 and 2 pH. The chlorine dosing system controls an SPST relay. Each alarm can be enabled or disabled.

pH control

3 pH concentration set points can be set: one control set point and two alarm set points. The pH control modes are user selectable; on/off or proportional dosing. The proportional dosing factor (1/Δ) is optional with Δ between 0.1 and 2 pH. The hysteresis value must be between 0.05 and 2.00 pH. The pH dosing system controls an SPST relay.

The pH, ORP and temperature alarms can be enabled or disabled, and two alarm levels can be set by the user for temperature. The alarm condition controls an SPDT relay. The fault alarm system activates a signal relay when process intervention is required. The fault alarm system controls an SPST relay.

Analog output

Two 4-20 mA or 0-20 mA outputs are available for connection to external devices such as chart recorders. The user can connect two dosing pumps via the 4-20 mA outputs for chlorine and acid/base dosing. The analog outputs are fully programmable and compatible with chlorine concentration, pH or temperature values. The analog output limits are selectable for each parameter.

Record data

Up to 3500 values can be stored (at least 7 days of recording at 3 minute intervals), which can be reviewed or downloaded. The recorded data records include the time, full information about the parameter values and the alarm status at that time.

Alarm and warning system

Through this system, the user can enable or disable low and high alarms for all parameters. The system also provides over-dose protection and alerts when the system is not working properly. The system will stop the process until the error is corrected by the user. The time is displayed on the main control panel and the reminder times are for “old standard”, “expired reagent”, and “expired SIM”. All these alerts are generated based on user settings.

Installation

Easy control with wall mount kit and supplied with chemicals.

(Note: Not for Bromine and Iodine)

Target options:

PCA310

Residual chlorine and total chlorine

PCA320

Residual Chlorine, Total Chlorine, pH and Temperature

PCA330

Residual Chlorine, Total Chlorine, pH, ORP and Temperature

 

SPECIFICATIONS

Chlorine

The scale

0.00 to 5.00 mg/L (ppm)

Resolution

0.01 mg/L (ppm)

Accuracy

± 8% or ±0.05 mg/L whichever is greater

EMC deviation

±0.05 mg/L (ppm)

Calibration

1 point

Sampling time

3 to 90 minutes

Dosage division

Proportional relay or 4-20mA output

Delta

Optional from 0.1 to 5 mg/L (ppm)

pH

The scale

0.00 to 14.00 pH

Resolution

0.01 pH

Accuracy

± 0.05 pH

EMC deviation

± 0.2 pH

Calibration

1; 2 point or in-line calibration

Sampling time

3 to 120 minutes

Dosage division

On/off or proportional, relay or 4-20mA output

Delta

Optional from 0.1 to 2 pH

Delay

Optional from 0.05 to 2 pH

ORP

The scale

0 to 2000 mV

Resolution

1 mV

Accuracy

± 1 mV

EMC deviation

± 10 mV

Temperature

The scale

5.0 to 75.0oC

Resolution

0.1oC

Accuracy

± 0.05oC

EMC deviation

± 0.5oC

Exit

4-20mA, 0-20mA

PC connection

RS485

Baud rate

1200, 2400, 4800, 9600 bps

Record data

Up to 3500 points

Alarm relay

SPDT 5A, 230V

Pump relay

SPDT 5A, 230V

Report an error

SPDT 5A, 230V

Pressure rate input

0.07 to 4 bar without external pressure regulator (for pressures exceeding 4 bar an external pressure regulator is required)

Flow rate

100 to 300 mL/min

Sample temperature

5 to 40°C (41 to 104°F)

Input/Output

12mm (1/2″) male NPT fitting

Connect drain

10mm (3/8″) barb

pH/t electrodeo

HI1005 (sold separately)

ORP electrode

HI2008 (sold separately)

Electricity supply

20 VA

Bark

NEMA-4X

Size

318 x 267 x 159 mm (12.5 x 10.5 x 6.25'')

Mass

5 kg (11 lb.) without reagent

Guarantee

12 months

Supply includes

PCA330 is supplied with 2 reagent bottles, 2 reagent caps, 1 DPD powder packet, tubing and instructions.

Probe purchased separately, not included

Note

PCA330-1: 115V

PCA330-2: 230V

Request a quote
SEMIKI MEASUREMENT EQUIPMENT COMPANY LIMITED
Headquarters:
  • Floor 12 – Tower A2, Viettel Building, 285 Cach Mang Thang Tam, Ward 12, District 10, Ho Chi Minh City, Vietnam
  • Sales@semiki.com
  • +84 979761016
  • Tax code 0313928935
Warranty, maintenance and repair center:
  • 10th Floor, Halo Building, 19-19/2A Ho Van Hue, Ward 9, Phu Nhuan District, HCMC
Representative office in Hanoi:
  • 9th Floor, 3D Building, No. 3 Duy Tan Street, Dich Vong Hau Ward, Cau Giay District, Hanoi City
© 2024 Semiki inc. All rights reserved.

Ask SEMIKI for a quote

Complete the form below to receive a quote from SEMIKI.
You need to enter all information fields below.