Technical specifications of testo 330 -2 LL exhaust gas meter
Measure differential pressure
Measuring range | ± 10,000 Pa |
Accuracy | ±0.3 Pa (0 to 9.99 Pa) ±1 digit ±3 % of measured value (10 to 10,000 Pa) ±1 digit |
O gas2
Measuring range | 0 to 21 Vol.% |
Accuracy | ±0.2 Vol.% |
Resolution | 0.1 Vol.% |
Response time | t₉₀ < 20 s |
CO gas measurement (H₂ compensation)
Parameter | |
---|---|
Measuring range | 0 to 8,000 ppm |
Accuracy | ±10 ppm or ±10 % of measured value (0 to 200 ppm) ±20 ppm or ±5 % of measured value (201 to 2,000 ppm) ±10 % of measured value (2001 to 8,000 ppm) |
Resolution | 1 ppm |
Response time | t₉₀ < 60 s |
Measure COlow gas
Parameter | |
---|---|
Measuring range | 0 to 500 ppm |
Accuracy | ±2 ppm (0 to 39.9 ppm) ±5 % of measured value (40 to 500 ppm) |
Resolution | 0.1 ppm |
Response time | t₉₀ < 40 s |
CO gas measurement (H₂ compensation), automatic dilution
Parameter | |
---|---|
Measuring range | 0 to 30,000 ppm |
Accuracy | ±100 ppm (0 to 1000 ppm) ±10 % of measured value (1001 to 30,000 ppm) |
Resolution | 1 ppm |
Measure NO gas
Parameter | |
---|---|
Measuring range | 0 to 3,000 ppm |
Accuracy | ±5 ppm (0 to 100 ppm) ±5 % of measured value (101 to 2,000 ppm) ±10 % of measured value (2001 to 3,000 ppm) |
Resolution | 1 ppm |
Response time t₉₀ | < 30 s |
Measure NOlow gas
Parameter | |
---|---|
Measuring range | 0 to 300 ppm |
Accuracy | ±2 ppm (0 to 39.9 ppm) ±5 % of measured value (40 to 300 ppm) |
Resolution | 0.1 ppm |
Response time t₉₀ | < 30 s |
Measure Draft pressure
Parameter | |
---|---|
Measuring range | -9.99 to +40 hPa |
Accuracy | ±0.02 hPa or ±5 % of measured value (-0.50 to +0.60 hPa) ±0.03 hPa (+0.61 to +3.00 hPa) ±1.5 % of measured value (+3.01 to +40.00 hPa) |
Resolution | 0.01 hPa |
.Temperature
Parameter | |
---|---|
Measuring range | -40 to +1200 °C |
Accuracy | ±0.5 °C (0 to +100.0 °C) ±0.5 % of measured value (remaining measuring range) |
Resolution | 0.1 °C (-40 to +999.9 °C) 1 °C (> +1000 °C) |
Combustion efficiency, Eta (calculated)
Parameter | |
---|---|
Measuring range | 0 to 120 % |
Resolution | 0,1 % |
Furnace efficiency (calculated)
Parameter | |
---|---|
Measuring range | 0 to 99.9 % |
Resolution | 0,1 % |
Calculate CO₂ (from O₂)
Parameter | |
---|---|
Measuring range | 0 to CO₂ max (display range) |
Accuracy | ±0.2 Vol.% |
Resolution | 0.1 Vol.% |
Response time t₉₀ | < 40 s |
Wind speed/air flow rate
Parameter | |
---|---|
Measuring range | 0.15 to 3 m/s |
Resolution | 0.1 m/s |
CO gas measurement (No H₂ compensation)
Parameter | |
---|---|
Measuring range | 0 to 4,000 ppm |
Accuracy | ±20 ppm (0 to 400 ppm) ±5 % of measured value (401 to 2,000 ppm) ±10 % of measured value (2,001 to 4,000 ppm) |
Resolution | 1 ppm |
Response time t₉₀ | < 60 s |
Ambient CO
Parameter | |
---|---|
Measuring range | 0 to 500 ppm |
Accuracy | ±5 ppm (0 to 100 ppm) ±5 % of measured value (> 100 ppm) |
Resolution | 1 ppm |
Response time | about. 35 s |
With CO probe
Ambient CO₂
Parameter | |
---|---|
Measuring range | 0 to 1 Vol.% 0 to 10,000 ppm |
Accuracy | ±50 ppm or ±2 % of measured value (0 to 5,000 ppm) ±100 ppm or ±3 % of measured value (5,001 to 10,000 ppm) |
Response time | about. 35 s |
With CO₂ probe
Leak measurement with flammable gases (via gas leak detection probe)
Parameter | |
---|---|
Measuring range | 0 to 10,000 ppm CH₄/ C₃H₈; Display range |
Accuracy | Signal optical display (LED) audible signal via buzzer |
Response time t₉₀ | < 2 s |
With air leak detection probe
Temperature (via pressure probe)
Parameter | |
---|---|
Measuring range | -40 to +1200 °C max. (dependent on probe) |
Accuracy | ±0.5 °C (-40 to 100 °C) ±0.5 % of measured value (remaining measuring range) + probe accuracy |
Resolution | 0.1 °C |
General specifications
Parameter | |
---|---|
Guarantee | Main equipment: 48 months; Probe: 48 months; Gas sensors (O₂, CO) 48 months; NO sensor 24 months; COlow sensor 24 months; Thermocouple 12 months; Rechargeable battery: 12 months |
Screen resolution | 240 x 320 pixels |
Display screen | Color screen |
Power supply | Rechargeable battery pack 3.7 V / 2.6 Ah; Mains unit 6 V / 1.2 A |
Maximum memory | 500,000 measurement values |