Semiki logo
HOTLINE: +84 979761016
0

Cart

No products in the cart.

Continue shopping
Home / Shop / Lab equipment / Hanna experimental measuring equipment / Mini device for analyzing acid concentration in fruit juice HI84532
Product portfolio

Mini device for analyzing acid concentration in fruit juice HI84532

Producer:
SKU: HANNA-HI84532 Category:

The HI84532 is a fast, convenient and easy-to-use titrator for the determination of acidity in fruit juices. The automatic titrator is equipped with a powerful algorithm and a dedicated pH electrode that allows the mini titrator to accurately determine the end point of the reaction. In addition, the HI84532 is equipped with a piston pump in dynamic pumping mode, which automatically adjusts the amount of reagent injected based on the change in sample potential during the titration, saving reaction time while ensuring reliable results. The HI84532 is supplied with all the necessary materials for determining the acidity of both the low and high ranges. Results are displayed as % citric acid, tartaric acid and malic acid. The chemicals supplied are premixed and packaged in a closed process including standards, reagents and pump calibration solutions, eliminating the need for additional equipment.

Piston pump with dynamic pump mode.

Standard solutions and reagents.

Capable of functioning as a dedicated benchtop pH meter

The acid content index is very important in determining the ripeness of fruit and the sourness of citrus fruits. In particular, ripeness is a very important parameter to determine the taste and the best method of preserving fruit. Many countries have their own standards for fruit (based on titratable acid content or the ratio of total soluble solids to titratable acid content (degrees Brix)) to protect the interests of consumers. Normally, unripe fruit will have a lower sugar/acid content ratio than ripe fruit.

The HI84532 will determine the acidity of fruit juice by neutralizing it with a strong base to a preset pH value, the titration value includes all naturally occurring acidic substances in fruit such as hydrogen ions, organic acids and acid salts. The acidity will be expressed in g/100 mL according to the acid with the highest content. Each type of fruit will have a main acid with the highest content. (refer to the table below for more information).

 

Glass pH electrode

The HI84532 is supplied with the HI1131B combined electrode, which features a spherical glass membrane with a large surface area, making it an excellent device for direct measurements or for use in acid-base titrations in the water industry.

pH calibration options

Since the instrument is set to titrate according to the AOAC standard method for fruits with an endpoint of pH 8.1, the HI84532 is set to optionally set the pH calibration point at 8.2 to achieve the most accurate results when titrating.

Good laboratory practice

HI84532 integrates GLP information storage function including electrode and pump calibration date, helping users understand the status of the instrument before use and can be more confident in the measurement results.

Piston pump.

The most important device of HI84532 is the piston pump. Using a motor to control the operation of the piston makes it easier to control the pumping process, and the amount of reagent pumped out is also more accurate. The pumping mode is set to dynamic, so the amount of reagent pumped in will change depending on the change in the potential in the sample after each pump, so when titrating, the machine will pump more reagent to shorten the time, when approaching the end point of the reaction, the amount of reagent will gradually decrease to ensure the accuracy of the measurement process.

 

Magnetic stirring feature.

A magnetic stirring device is integrated into the machine, automatically maintaining a speed of 600 rpm, making it easy for the device to titrate highly viscous samples.

 

On-screen features

Easy-to-use interface: HI84532 has a user-friendly design, sharp keypad, large screen for easy use. In addition, the device is installed with a tutorial software, when activated, will guide the user on how to use the device step by step. During use, if there are any questions, the user can press the HELP key for information support.

Warning sign: users will be alerted in case errors occur when using the machine (eg: reagent volume exceeds the allowable limit..)

Graph mode: the mini titrator displays data throughout the titration, including real-time graphs.

Logging on Demand: The HI84532 allows for logging of up to 400 samples: 200 titration results and 200 pH/mV measurements. These data can be stored to a USB drive or transferred to a PC via the USB port.

As a pH/mV meter: in addition to being a titrator, the HI84532 can also be used as a benchtop pH meter, with automatic temperature compensation, automatic buffer recognition, up to 3-point calibration, ±0.01pH accuracy, and GLP data storage.

CAL Check feature. The machine will automatically analyze the electrode response signal during the calibration process and notify the user of problems that may affect the measurement results such as contaminated buffer solution, dirty electrode or cracking… the CAL Check feature not only ensures the accuracy of pH measurements but also helps ensure accuracy in determining the end point of the titration process.

YouTube video

SPECIFICATIONS

The scale

Low ladder:

g/100 mL as citric acid: 0.10 to 2.00% CA;

g/100 mL as tartaric acid: 0.11 – 2.35% TA;

g/100 mL as malic acid: 0.10 – 2.09% MA;

Height:

g/100 mL as citric acid: 1.00 – 10.00% CA;

g/100 mL as tartaric acid: 1.17 to 11.72% TA;

g/100 mL as malic acid: 1.05 – 10.47% MA

Resolution

0.01%

Accuracy (@25ºC/77ºF)

±0.02% CA or 3% measured value, take the larger value

Sample volume

5mL

Method

acid-base titration

Rule

End point pH 8.1

Pump speed

10 mL/min

Stirring speed

600rpm

pH scale

-2.0 to 16.0pH; -2.00 to 16.00 pH

pH resolution

0.1 pH / 0.01 pH

pH accuracy(@25ºC/77ºF)

±0.01 pH

pH calibration

Maximum 3 points with the following buffers: 4.01, 7.01, 8.20, 10.01

Heat compensation

Manual or automatic

mV scale

-2000.0 to 2000.0 mV

mV resolution

0.1 mV

mV accuracy (@25ºC/77ºF)

±1.0 mV

Temperature scale

-20.0 to 120.0°C; -4.0 to 248.0°F; 253.2 to 393.2 K

Temperature resolution

0.1°C; 0.1°F; 0.1K

Temperature accuracy (@25ºC/77ºF)

±0.4°C; ±0.8°F; ±0.4 K

Save the template

Maximum 400 (200 titrations, 200 pH/mV measurements)

Electrode

HI1131B glass pH electrode, BNC port, 1m cable

Temperature electrode

HI7662-T, 1m cable

Operating environment

0 – 50°C (32 – 122°F); Max RH 95% non-condensing

Source

12 VDC adapter

Size

235 x 200 x 150 mm 

Mass

1.9kg 

Guarantee

12 months for body and 6 months for electrodes

Supply includes

HI84532-02 (230V) is supplied with HI1131B pH electrode, HI7662-T temperature electrode, titration solution set, HI84532-55 calibration solution (100 mL), HI7082 electrode fill solution (30 mL), 100 mL plastic beaker (2), pump valve, 5 mL syringe, 1 mL plastic pipette, tubing, magnetic hook, 12 VDC adapter, instructions, and quality certificate.

Request a quote
SEMIKI MEASUREMENT EQUIPMENT COMPANY LIMITED
Headquarters:
  • Floor 12 – Tower A2, Viettel Building, 285 Cach Mang Thang Tam, Ward 12, District 10, Ho Chi Minh City, Vietnam
  • Sales@semiki.com
  • +84 979761016
  • Tax code 0313928935
Warranty, maintenance and repair center:
  • 10th Floor, Halo Building, 19-19/2A Ho Van Hue, Ward 9, Phu Nhuan District, HCMC
Representative office in Hanoi:
  • 9th Floor, 3D Building, No. 3 Duy Tan Street, Dich Vong Hau Ward, Cau Giay District, Hanoi City
© 2024 Semiki inc. All rights reserved.

Ask SEMIKI for a quote

Complete the form below to receive a quote from SEMIKI.
You need to enter all information fields below.